@article{YPJ1209, author = {Indri Trifiantari and Suryono Suryono}, title = {RANCANG BANGUN ALAT PENGUKUR DISTRIBUSI KUAT PENERANGAN MULTI TITIK MENGGUNAKAN SISTEM KOMUNIKASI RS485}, journal = {Youngster Physics Journal}, volume = {1}, number = {1}, year = {2012}, keywords = {}, abstract = { ABSTRACT Meassuring illuminance system was created used multipoint m odel . The used light sensor wa s LDR that created used voltage devider model . AVR ATMega8535 microcontroller wa s used as a data acquisition device and a serial data transmission to a computer. O utput of sensor wa s inserted into the microcontroller ADC input at ADC0 pin . Measured data of the LDR sensor w ou l d be processed by AVR ATMega 8535 microcontroller , then sent to a computer using a serial RS485 protocol. Th e data was interface d with Borland Delphi7 and s av ed o n Microsoft Excel. From this research, it was found the measurement from 4 terminal systems which each terminal sistem consist of one measurement point. From the four test systems performed by standard luxmeter obtained equipment precision of device that is made between R = 94.84%, until R = 99.05%. Key words: Illumination , LDR, RS485 communication system }, issn = {2302-7371}, url = {https://ejournal3.undip.ac.id/index.php/bfd/article/view/1209} }
Refworks Citation Data :
ABSTRACT
Meassuring illuminance system was created used multipoint model. The used light sensor was LDR that created used voltage devider model . AVR ATMega8535 microcontroller was used as a data acquisition device and a serial data transmission to a computer. Output of sensor was inserted into the microcontroller ADC input at ADC0 pin. Measured data of the LDR sensor would be processed by AVR ATMega 8535 microcontroller, then sent to a computer using a serial RS485 protocol. The data was interfaced with Borland Delphi7 and saved on Microsoft Excel.
From this research, it was found the measurement from 4 terminal systems which each terminal sistem consist of one measurement point. From the four test systems performed by standard luxmeter obtained equipment precision of device that is made between R = 94.84%, until R = 99.05%.
Last update: